10 Start Screen

Once the stage and advance are fully retracted, the Start Screen is displayed.

The Start Screen gives access to the tools required to prepare the slicer for the next tissue sample.FigureThe Start screen can be entered either from a fresh power up or from the Menu screen, by pressing the FigureStart button.

The Stage and Advance axis of the instrument are moved to the "Load" position (the stage and advance are fully retracted).

The screen gives information on blade usage. If a blade is changed, then the 'Blade' information will appear on a red background.

From the Start screen, the following options can be found:

  1. Slice: Select Figure to Load the tissue bath to the home position and begin slicing. A tissue bath must be fitted to proceed. See section 13 to position the blade ready for the first slice.
  2. Change: Select the change Figure button when a blade change is required. This option allows a reset and adjustment (if required) of the blade change timer. It also allows for a selection of blade type being used; this is important as the instrument uses this for the Opti-Cal routine. Refer to section 8 when changing the blade.
  3. Calibrate: Select Figure to set the blade edge parallel to the cutting plane for optimum slice quality. See Blade Alignment in section 11.2.
  4. Change Tissue Mount type: touch the image corresponding to the correct Tissue Mount Figure Figure Figure to select. This allows the instrument to adjust the slice datums according to the different dimensions of the Tissue Mount. Default is set to "Standard."
  5. Custom: Select Figure load a different Custom Setup (see Section 15). The current setup is displayed on screen.
  6. Menu: Select Figureto access the Menu screen (see Section 16).

Proceed to section 11.

Yellow Warning If this is the first time using this instrument, you will need to select, calibrate the blade and choose the tissue mount. CUSTOM will be irrelevant for now as the 9000SMZ only has factory defaults set.