14 Tissue Specimen Slicing

When the blade is optimized and positioned correctly relative to tissue specimen, the slicing can proceed.
Figure">Figure
The slice screen & Keypad

14.1 Slice parameters

Slice parameters are all displayed on the slice screen and can be updated by pressing the desired parameter button (Cycles, Slice Parameters and Section) or Figure from the menu. This will open the Slice Parameters Screen. Slice parameters can be stored as part of the Custom Setup as described in Section 15, or by holding the Figure Slice button on exit.

 

Figure

Section
The Figure Section selection button, allows the sequential selection of three section thicknesses (denoted by the number 1-3 adjacent to the ruler symbol); each slice thickness option may be edited within a range of 001µm to 999µm using the number pad. See Section 14.5.

Cycles
See Section 14.6

Frequency/Amplitude Matrix
Within the Parameters Screen selecting the desired frequency will highlight the available amplitudes in cyan. Alternatively selecting the desired amplitude first will highlight the available frequencies in light blue. The selected frequency and amplitude will be highlighted in violet.FigureFigure Note: for the higher frequencies, only the lower amplitudes are possible, see the table below.

Figure Select the Figure Slice button to return to the Slice Screen.

14.2 First Pass

Fine tuning of the blade position relative to the mounted tissue specimen may be achieved prior to taking a slice by operating the Blade Advance Rotary Dial. The stage height can be adjusted using the up and down buttons on the keypad. This will move the stage respectively and the Stage position will be updated on the screen. Pressing the Figure Step move button will change the adjustment value. Note: For the first pass, the section value will not be affected by this adjustment.

14.3 Return to start

Selecting the Figure return button after a slice, the blade advance will retract to the selected Start position. As part of the return move, the stage is retracted slightly to avoid rubbing across the specimen. When the Start position is reached the stage will be incremented by the selected specimen thickness, ready for the next slice.

14.4 Slice Mode

Two Slice Window modes are available and can be selected by toggling the Slice Window mode button between Figure Automatic slice window/profile and Figure Manual slice window. Automatic is the default setting, though this may be changed as part of the Custom Setup (see Section 15). The following list shows how the different slice modes operate:

  1. Figure Automatic Window Procedure
    FigureAuto-Mark Button
    If the Auto-Mark button is selected Figure the most recent slice image will be recorded after the slice has been taken and the return button pressed. If the button is not enabled Figure , any previously stored image will be available.

    FigureFigureFigureAuto-Window/Profile Button
    The Auto-Window /Profile button toggles between three modes, Manual Slice Figure , Window Slice Figure and Profile Slice Figure. Window Slice and Profile Slice are only available once an image has been recorded, usually by taking the first slice.

    Figure + FigureAutomatic Mark and Recording an Image (Default)Figure
    When first entering the slice screen, this mode is set by default. Automatic Slice Mode facilitates the “Auto-marking” of Start and Finish points by recording them during an active slice.

    Press the Figure Slice button on the keypad. As the slice is being taken, the user has full control of the blade advance speed and the ability to pause the Advance via the Rotary Dial. Figure
    During the slice, the advance can be paused (pressing the Rotary Dial) or the blade can be stopped and re-started by pressing the Slice Key Figure.

    When the blade has advanced through the tissue, press the Figure Slice Key to stop the slice.
    The touchscreen will prompt “Image will create on return”. If the image is not required, press the FigureAuto-Mark button to de-select , the touchscreen will then prompt “Set mark to create image on return”.

    Pressing the Figure Return Key will move the Blade, Advance and Stage position to the start point of the next slice. When the Figure Return Key is pressed the image either be kept or discarded, depending on the status of the Auto-Mark Button.
    Figure If the Figure Auto-Mark button is selected to create an “image”, the recorded “image” will be stored from this slice and be available for subsequent slices. Note: The “image” of the slice contains the slice’s FigureStart and FigureFinish points, plus the variations in the Advance speed (Profile) made by the user during the slice.
    Figure
    Figure + Figure + Figure Automatic Window Slice
    This mode uses the Figure Start and Figure Finish points of the most recently stored image. An image must be available. The advance speed between these points is controlled by the user. Select the profile button Figure until the Figure "window" icon” appears in blue.
    Note:
    Ensure that Figure Auto-Mark button is deselected if you do not wish to overwrite the slice image.
    If required, select the desired Figure Cycle option, refer to Section 14.6.
    Figure
    This image shows an Auto Window slice with the Cycles option selected.
    Press the Figure Slice Key. Figure This image shows an Auto Window slice with the Cycles option selected. The Blade & Advance are active.

    Adjust Advance speed as desired during slicing via the Rotary Dial.
    Subsequent slices may be taken until the Limit of Stage travel has been reached.

    Figure + Figure + Figure Automatic Profile Slice Procedure
    From the most recently stored image, Figure Start, Figure Finish points and the Advance speed profile across the slice may be used for subsequent slices.
    Toggle the Figure Window/Profile button until the Figure Profile button is display.
    When the slicing is "active" the user can pause the Advance by momentarily pushing on the Rotary Dial, but will be unable to adjust the Advance speed. To return to manual control during a slice, toggle the Figure window/profile button to deselect. The speed Advance speed is now controlled by the Rotary Dial.
    Select the Figure Profile (window/profile) button (this will highlight in blue).
    Note: Ensure that Figure Auto-Mark button is deselected if you do not wish to overwrite the slice image. Figure This image shows an Auto Profile slice with the Cycles option de-selected.
    Press the Figure Slice button on the keypad. Figure This image shows an Auto Profile slice with the Cycles option de-selected. The Blade and Advance are active.
    Repeat slices may be taken until the Limit of Stage travel has been reached.
  2. Figure Manual Slice Procedure
    In this mode it allows you to set FigureStart and Figure finish limits manually when a slice is taken.
    Before slicing, a Figure Start limit can be created by pressing and holding the Figure Start limit button. Momentary pressing this button toggles the set limit on and off (ON = Figure).
    Press the Figure slice button on the keypad to take a slice. Figure When the slice has been taken it is possible to mark a Figure Finish point by pressing and holding the Figure Finish button. Momentary pressing this button toggles the set limit on and off (ON = Figure).
    Press the Figure Return button on the Keypad to return to the Start Limit. If a limit is not set, then the blade will return to the blade advance home position.
    If the Finish limit has been set, subsequent slices will be Windowed. It is possible to select a Figure cycle mode (see section 14.6). The home position will be used as the start limit if none is set/selected.
    Repeat slices may be taken until the Limit of Stage travel has been reached.

14.5 Section Thickness.

There are three preset Section thickness available, they are selected by pressing the Figure section selector button. The preset values can be set in the Slice Parameters Screen (see section 14.1).

The section thickness will automatically be added to the stage travel after a Figure Return operation. Making changes after this will automatically readjust the stage height to suit.

There are three ways to make a quick change (Note: this is not possible on the first pass). If the pre-set value is not used, changing the value will move the section selection to #.

14.6 Repeat Cycles, Stop at Start and Stop at Finish.

It is possible to enable a fully automatic repeat slice operation, a Stop at start operation and a Stop at Finish operation. They are available for both Automatic and Manual slice modes. Figure + Figure / Figure + Figure Window or Figure + Figure Profile repeat must have been selected to enable these modes.

Mode selection

The selected mode is indicated on the Figure Cycle mode button. Pressing this button will enable the function Figure. To change the mode, Press the large cycles display to enter the Slice Parameters screen (see section 14.1). Toggle the Cycles icon here to select the mode and enter the number of cycles for the repeat function using the numpad. These parameters can be saved to the Custom Setup (see section 15).

FigureRepeat Cycles

With this mode enabled the slicer will continually repeat the windowed or profile slice operation for the number of cycles indicated.

FigureStop at Start

With this mode enabled, on pressing the slice button the instrument will take either a window slice or a Profile slice. At the finish point, the slice will automatically return to Start ready for the next slice operation.

FigureStop at finish

With this mode enabled, on pressing the slice button the instrument will take either a window slice or a Profile slice. At the finish point, the slice will stop. Pressing slice at this point, the instrument will Return to the Start point and begin taking the next slice. Pressing Return when at the finish point will make a return move only.